General Questions

  • Qid® is a revolutionary, FDA cleared, first of its kind technology that provides the patient with an electronic serial number from within their body for complete control and verification of the breast implants. This ensures extended safety and wellbeing for the patient and precise data in the unlikely event of a recall or an adverse event.
    Following surgery, the patient can return to the surgeon´s to have her implant(s) verified and registered via the Motiva Qid® Handheld Reader. The 15-digit number delivered to the reader corresponds with a secure, online database that can be accessed via the internet and by the surgeon and the patient only.

    Learn more
  • The start button is located on the back of the reader. To power the unit on, simply push on the symbol located on the back panel by the words “press here” and then release.
    The display screen will show Motiva Qid® Reader “Reading” and the reader will emit a single beep. If no valid Electronic Serial Number (ESN) is found within 20 seconds, then the display window will indicate “No Device Found” and the reader will again emit a single beep. If there is no activity within 40 seconds, the reader will automatically turn off.
    The outer surface of the Motiva Qid® Reader is the reading area. When a valid ESN is found, the reader will emit a double beep sound and the complete ESN number will be shown on the display screen.

    Learn more
  • When an ESN is located, you will hear a beep, followed by a 15-digit numeric sequence shown on the display screen. If no MT is found, try reading again moving slowly in different directions and varying the angles of the reader on the breast area. If no ESN is still found, re-test the Motiva Qid® Reader with the test ESN provided.
    Further operation instructions for the Motiva Qid® Reader are located in the User Manual, which can be downloaded from the flash drive included with the reader.

    Learn more
  • The Motiva Qid® Reader reader should never be submersed in liquid or placed in any sterilization equipment. It can be easily wiped down with a Steriwipe® or by moistening a cloth with a disinfecting solution.

    Learn more
  • When the battery is running low on the Motiva Qid® Reader, the following message will appear on the display screen - “Battery Low”. To recharge the battery, simply plug the USB cable into the back of the reader and into an available USB port.
    A fully discharged battery will need to charge for up to three and half hours in order to be fully recharged. During the charging cycle, the display will switch off. To check if the battery has finished charging, simply press the start button and the display screen will inform you if it is still “Charging” or “Charged”.

    Learn more
    • The device is ROHS compliant and meets ISO 11784 and 11785
    • Dimensions are 135mm diameter x 33mm (5.315 inches diameter x 1.299 inches deep)
    • Weight is 70g (2.4962 ounces)
    • Reads per charge are 8 second scans x 1000. The battery capacity can vary in time with normal use
    • Charge time is 3.5 Hours
    • Operating temperature is 0°C to +50°C (32° F to +122° F)
    • Software - Windows 98 or later
    • Warranty - 1 Year manufacturer’s warranty against defects in materials or workmanship that prevent its use
    • Atraumatic handling minimizes tissue damage and increases graft viability.
     
    Learn more
See all 7 articles

Contact Motiva®!

Have another question? Send us a message!